| Home | E-Submission | Sitemap | Contact Us |  
Clinical and Experimental Otorhinolaryngology > Volume 13(2); 2020 > Article
Yeon, Sohn, Seo, Kim, Eun, Park, and Yun: Diagnostic Performance of a Combination of Shear Wave Elastography and B-Mode Ultrasonography in Differentiating Benign From Malignant Thyroid Nodules

Abstract

Objectives

This study was conducted to compare clinicopathologic and radiologic factors between benign and malignant thyroid nodules and to evaluate the diagnostic performance of shear wave elastography (SWE) combined with B-mode ultrasonography (US) in differentiating malignant from benign thyroid nodules.

Methods

This retrospective study included 92 consecutive patients with 95 thyroid nodules examined on B-mode US and SWE before US-guided fine-needle aspiration biopsy or surgical excision. B-mode US findings (composition, echogenicity, margin, shape, and calcification) and SWE elasticity parameters (maximum [Emax], mean, minimum, and nodule-to-normal parenchymal ratio of elasticity) were reviewed and compared between benign and malignant thyroid nodules. The diagnostic performance of B-mode US and SWE for predicting malignant thyroid nodules was analyzed. The optimal cutoff values of elasticity parameters for identifying malignancy were determined. Diagnostic performance was compared between B-mode US only, SWE only, and the combination of B-mode US with SWE.

Results

On multivariate logistic regression analysis, age (odds ratio [OR], 0.90; P=0.028), a taller-than-wide shape (OR, 11.3; P=0.040), the presence of calcifications (OR, 15.0; P=0.021), and Emax (OR, 1.22; P=0.021) were independent predictors of malignancy in thyroid nodules. The combined use of B-mode US findings and SWE yielded improvements in sensitivity, the positive predictive value, the negative predictive value, and accuracy compared with the use of B-mode US findings only, but with no statistical significance.

Conclusion

When SWE was combined with B-mode US, the diagnostic performance was better than when only B-mode US was used, although the difference was not statistically significant.

INTRODUCTION

Ultrasonography (US) is the first-line and most preferred examination method for thyroid nodules. Conventional US can be used to detect nodules, and it also provides information on the composition, size, and number of nodules, as well as their anatomical relationships with surrounding structures. US findings that have been reported to be suggestive of malignant thyroid nodules include solid composition, hypoechogenicity or marked hypoechogenicity, a microlobulated or irregular margin, mixed calcification or microcalcifications, and a taller-than-wide shape [1,2]. However, a considerable number of US findings overlap between benign and malignant thyroid nodules, making it difficult to identify thyroid malignancies based on US imaging alone [3,4]. Fine-needle aspiration (FNA) can play an important role in differentiating thyroid nodules because it has a high sensitivity and specificity. However, the diagnostic performance of FNA depends on the operator’s experience, it is an invasive method, and it yields a high proportion of indeterminate (10%–20%) or nondiagnostic (10%–15%) results [5,6]. The recently published World FedEration for Ultrasound in Medicine and Biology guideline recommends US elastography as an additional modality in thyroid nodules to support surgical excision and periodic observation [7]. US elastography is a new sonographic imaging technology that estimates tissue hardness and helps to discriminate benign from malignant nodules. The basic concept of US elastography was first developed in 1991 by Ophir et al. [8]. Most malignant nodules have abnormally firm stroma because of the presence of myofibroblasts, which make it possible to differentiate thyroid cancer from benign nodules using elastography. Nell et al. [9] conducted a systematic review and meta-analysis of 20 studies. Their analysis demonstrated that unnecessary invasive diagnostic procedures such as FNA could be avoided in completely soft (Asteria elastography 1) thyroid nodules [10] with a 99% negative predictive value.
There are two kinds of US elastography: strain elastography and shear wave elastography (SWE) [11]. The limitations of strain elastography include difficulties with the quantitative estimation of tissue hardness and its low reliability as a result of operator dependency. In contrast, SWE uses transversely oriented shear waves created by mechanical vibration forces. The shear waves propagate faster in stiff tissue than in soft tissue, thus enabling a quantitative approach using color maps and elasticity indices including Emax, Emean, Emin, and Eratio (with units of kilopascals [kPa]) [11]. This new technique has made it possible to evaluate tissue stiffness in a manner that is more quantitative, more reproducible, and less operator-dependent [12,13]. Some comparative studies found that SWE had lower interobserver variability than strain elastography; thus, it was preferred to real-time elastography and regarded as a promising tool with higher diagnostic performance and the ability to provide quantitative results for differentiating thyroid nodules [14].
Until now, the majority of US elastography studies for predicting thyroid malignancies have investigated strain elastography [15-17]. To the best of our knowledge, only a few studies have investigated the use of SWE to identify thyroid malignancies, and those studies showed varying results in terms of diagnostic performance and the optimal cutoff values of elasticity parameters [18-21]. Therefore, the aim of this study was to compare clinicopathologic and radiologic factors between benign and malignant thyroid nodules and to evaluate and verify the diagnostic performance of a combination of SWE and B-mode US in distinguishing histologically proven malignant from benign thyroid nodules.

MATERIALS AND METHODS

Patients

This retrospective study of thyroid nodule analysis had institutional review board approval, and the need to obtain patient informed consent for the study was waived. From March 2016 to July 2017, 224 neck lesions from 214 consecutive patients underwent neck sonography using B-mode US and SWE before US-guided FNA (US-FNA) biopsy or surgical removal. Exclusion criteria were as follows: (1) non-thyroid lesion such as parotid gland, cervical lymph node (113 lesions in 106 patients), (2) unavailable for SWE imaging or elasticity parameters (one lesion in one patient) and (3) cytological results of non-diagnostic lesion or atypia of undetermined significance without surgical confirmation (15 lesions in 15 patients). Finally, 95 thyroid lesions in 92 patients were included in this study and finally diagnosed as benign and malignant by cytopathological confirmation as reference standard (US-FNA, n=55; surgical excision, n=40). Patients consisted of 23 men and 69 women with a mean age of 52.12 years (range, 22 to 82 years). Fifty-four were benign thyroid nodules and 41 thyroid nodules were malignant.

US examination

All US examination including B-mode US and SWE scans were conducted with a 4-15 MHz linear transducer and a real-time US system (Aixplorer, SuperSonic Imagine, Aix en Provence, France) by two radiologists (YMS and MS) blinded to the result of histopathology with 7–11 years of experience in thyroid US and 6 years of elastography experience. Patients lied on the bed in a supine position with neck extension, the thyroid nodules and the neck were scanned carefully.
Each thyroid lesion was characterized according to B-mode US findings: composition (solid, mixed), echogenicity (marked hypoechoic, hypoechoic, isoechoic, mixed echoic), margin (irregular or microlobulated, circumscribed), calcifications (presence, absence) and shape (taller-than-wide, wider-than-taller). B-mode US characteristics for suspicious malignant findings were as follows; solid, marked hypoechogenicity or hypoechogenicity, irregular margin, presence of calcification, or taller-than-wide shape. The American Thyroid Association (ATA)’s guidelines for the management of thyroid nodule suggested five categories based on sonographic patterns [22]. (1) Nodules with the high suspicion of malignancy pattern, including solid hypoechoic nodule or solid hypoechoic component of a partially cystic nodule with one or more of the following US features: irregular margin, microcalcification, taller-than-wide shape, rim calcifications with small extrusive soft tissue component and evidence of extrathyroidal extension (Fig. 1). (2) Nodules with the intermediate suspicion of malignancy pattern, including hypoechoic solid nodule with smooth margins without microcalcifications, extrathyroidal extension, or taller than wide shape (Fig. 2). (3) Nodules with the low suspicion of malignancy pattern, including hyperechoic or isoechoic solid nodule, or partially cystic nodule with eccentric solid areas, without microcalcification, irregular margin or extrathyroidal extension, or taller than wide shape (Fig. 3). (4) Nodules with the very low suspicion pattern include spongiform or partially cystic nodules without any of the US features described in low, intermediate, or high suspicion patterns. (5) Benign pattern includes purely cystic nodules (no solid component). We used this ATA’s guidelines, however, very low suspicion and benign patterns were not included in this study because these nodules were not referred to US-FNA.
SWE imaging was obtained on longitudinal sections in freehand technique. The probe was placed gently in the patient’s neck without external compression. The region of interest (ROI) (Q-box; SuperSonic Imagine) of the system was placed in the thyroid lesion with surrounding normal tissue, which showed translucent colored area of tissue hardness overlaid B-mode image with a range from blue to red. The lowest hardness was presented blue and the highest hardness was presented red (0-180 kPa). Fixed 2×2-mm ROIs were located over the hardest part of the lesion with optimal image showing color homogeneity on the screen. Secondary same sized ROI was located in the normal thyroid tissue. The SWE system analyzed elasticity parameters of Emean (mean value of the ROI located in the hardest region), Emax (maximum value of the ROI located in the hardest region), and Emin (minimum value of the ROI located in the hardest region), and the Eratio (ratio of mean stiffness for the lesion-to-normal parenchyma) was calculated automatically.

Statistical analysis

To evaluate the difference between benign and malignant nodule, either the independent t-test or Mann-Whitney U-test was used for the continuous variables and Pearson’s chi-square test or Fisher’s exact test was performed for the nominal variables. To assess the independent factor for discriminating thyroid malignancy from benign thyroid nodule, the multivariate logistic regression analysis was applied. Adjusted odds ratios (ORs) and 95% confidence intervals (CIs) were generated from the multivariate analyses.
To generate diagnostic performance and the cutoff values of the B-mode US and elastographic parameters for differentiation malignant thyroid nodule from benign nodule, a receiver operating characteristic (ROC) curve analysis was performed and areas under the ROC curves (AUCs) were analyzed. Therefore, diagnostic accuracies, sensitivities, and specificities were calculated in B-mode US alone, elastography alone, and combined B-mode US with elastography using the cutoff values of the B-mode US and elastographic parameters. All statistical analyses were performed using MedCalc software ver. 12.3.0 (Mariakerke, Belgium). The P-values <0.05 were considered statistically significant.

RESULTS

Demographic, clinical, and sonographic characteristics

The 92 subjects included 23 men (mean age, 56.9 years; age range, 30.0 to 73.0 years) and 69 women (mean age, 51.3 years; age range, 22.0 to 76.0 years). Their characteristics are summarized in Table 1. Fifty-four thyroid nodules were benign and 41 were malignant. The thyroid nodules confirmed to be malignant included 39 papillary thyroid carcinomas, one medullary carcinoma, and one anaplastic carcinoma.
The patients with malignant thyroid nodules were significantly younger than those with benign nodules (P=0.009). The mean size of malignant nodules on US (12.0±7.3 mm) was not significantly different from that of benign lesions (15.7±10.2 mm; P=0.120) There was no significant difference according to sex or mean US nodule size between benign and malignant nodules. Suspicious US findings (solid composition, hypoechogenicity, an irregular margin, presence of calcifications, and taller-than-wide shape) and the US category of high suspicion were significantly more frequent in malignant nodules than in benign nodules (all P<0.001). Except Eratio, the mean values of all elasticity parameters were significantly higher in malignant thyroid nodules than in benign nodules (P<0.05) (Table 1).

Predictors of thyroid malignancy

On multivariate logistic regression analysis, age (OR, 0.90; P=0.028), a taller-than-wide shape (OR, 11.3; P=0.040), the presence of calcifications (OR, 15.0; P=0.021), and Emax (OR, 1.22; P=0.021) were found to be independent risk factors for malignancy in thyroid nodules (Table 2).

Diagnostic performance of each US feature, US category, and elasticity parameter and optimal cutoff values on SWE for predicting thyroid malignancy

Hypoechogenicity showed the highest AUC (0.838) of the Bmode US features. Among the quantitative SWE parameters, Emax showed a somewhat higher AUC (0.633) than the other elasticity parameters, including Emean, Emin, and Eratio, with a sensitivity of 53.66% and a specificity of 70.37%. The optimal threshold values of the elasticity parameters for predicting malignancy in thyroid nodules were as follows: Emean, 54.9 kPa; Emax, 47.1 kPa; Emean, 40.4 kPa; and Eratio, 1.94. The US category of high suspicion showed a sensitivity of 95.1%, a specificity of 85.2%, and an AUC of 0.913, which were all higher than the corresponding values of the SWE parameters (Table 3).

Diagnostic performance of combinations of SWE parameters with B-mode US findings

Combining the US category of high suspicion with the elasticity parameters generally yielded better diagnostic performance than using only the US category. Combining the US category of high suspicion with Emax resulted in a higher sensitivity than using the US category alone, as well as a negative predictive value of 100%. Combining the US category of high suspicion with Emin demonstrated a higher specificity and positive predictive value than using the US category alone (88.89% and 86.96%, respectively). A higher accuracy (92.63%) resulted from combining the US category with Emin. However, combining the US category of high suspicion with SWE parameters did not yield significant improvements in diagnostic performance compared with the use of the US category alone (Table 4).

DISCUSSION

Our study revealed that patients with malignant thyroid nodules were significantly younger (mean age, 47.6 years) than those with benign nodules (mean age, 55.5 years). This is consistent with the results of a previous large-scale study of the use of SWE to detect cancer in thyroid nodules that found the mean age of patients with a thyroid malignancy to be 42.9 years, while the mean age of those with benign nodules was 52.2 years [23]. When assessing indeterminate or suspicious thyroid nodules, a more careful evaluation is required if the patient is relatively young.
We investigated B-mode US finding features that could be useful for differentiating malignant from benign thyroid nodules, and found that solid composition, hypoechogenicity, an irregular/microlobulated margin, a taller-than-wide shape, and the presence of calcifications were significantly more common in malignant thyroid nodules than in benign thyroid nodules. These results are consistent with those of previous studies on US findings suggestive of malignancy in thyroid nodules [24-26]. In the present study, we compared and evaluated the thyroid nodules using SWE. All elasticity parameters except Eratio (Emax, Emean, and Emin) on SWE were significantly higher for malignant thyroid nodules than for benign thyroid nodules. This observed relationship between elasticity and the likelihood of thyroid malignancy is consistent with several prior publications analyzing the use of SWE [11,18-21,23].
Among the elasticity parameters, only Emax was found to be an independent risk factor for malignancy in thyroid nodules in the multivariate logistic regression analysis (OR, 1.22; 95% CI, 1.20 to 1.57, P=0.021). Emax had the best diagnostic performance, with a sensitivity, specificity, and AUC of 53.7%, 70.4%, and 0.633, respectively. In concordance with the present results, Wang et al. [27] also reported, in a study of 322 thyroid nodules, that Emax was the only independently significant SWE imaging parameter, with a sensitivity of 59.40%, a specificity of 89.8%, and an AUC of 0.796 for differentiating between malignant and benign thyroid nodules.
Several studies have reported a somewhat variable range of values of SWE imaging parameters that could be used to predict thyroid malignancies, especially regarding the optimal cutoff value of Emax. In our study, in which 43% of the nodules were malignant, the optimal cutoff value of Emax was 47.1 kPa, while those of Emean and Emin were 54.9 kPa and 40.4 kPa, respectively. A prior study reported similar results to ours, demonstrating that the optimal cutoff value of Emax was 47.0 kPa for predicting malignancy in 322 nodules, of which 33% were malignant [27]. However, Liu et al. [21], in an analysis of 313 nodules, of which 194 were malignant, reported that the optimal cutoff value of Emax was 51.9 kPa, with a sensitivity of 81.4% and a specificity of 83.2%. This may reflect the higher proportion of malignant nodules than in our study, because malignant nodules show higher SWE values than benign nodules. In contrast, Veyrieres et al. [19] analyzed 297 nodules, of which only 12% were malignant, and reported a high threshold cutoff value of Emax (66 kPa) with high sensitivity (80.0%), specificity (90.5%), and negative predictive value (97.1%), but with a relatively low positive predictive value (52.8%). Thyroid malignancies with a maximum diameter equal to or less than 1 cm showed significantly lower SWE values than nodules larger than 1 cm, probably because larger nodules might be subjected to higher precompression due to their convex bulging through the skin [28]. While the mean diameter of the thyroid malignancies studied by Veyrieres et al. [19] was 25.0±13.8 mm, the mean diameter was 15.7±10.2 mm in our study. As discussed above, the optimal cutoff values of SWE parameters vary considerably due to differences in the number of enrolled patients, the proportion of malignant thyroid nodules, and nodule size. Furthermore, several other characteristics of nodules, such as the presence of calcifications or cystic changes and their position, could affect the results [28]. Therefore, it is necessary to further assess the optimal cutoff values for predicting malignant thyroid nodules in a general screening population in a future study.
A previous study including 313 thyroid nodules reported that combining B-mode US findings with SWE parameters improved diagnostic performance for detecting thyroid malignancy, with the sensitivity and negative predictive value increasing from 89.7% and 84.0% to 94.6% and 90.0%, respectively [21]. However, the specificity and positive predictive value slightly decreased when B-mode US findings and SWE were combined. Another previous large-scale study involving 476 thyroid nodules showed similar results. The combination of B-mode US findings and elasticity parameters significantly increased the sensitivity, negative predictive value, and AUC from 92.9%, 68.7%, and 0.769 to 95.0%, 74.0%, and 0.834, respectively, but with slight decreases in the specificity and positive predictive value [18]. In our study, we observed slight improvements in all parameters of diagnostic performance as a result of combining the B-mode US category of high suspicion with each elasticity parameter, compared with using only the B-mode US category. It was especially meaningful that the sensitivity and negative predictive value increased from 95.1% and 95.8%, respectively, when using only the B-mode US category, to 100% and 100%, respectively, when combining the B-mode US category with Emax. Screening thyroid US requires a high sensitivity and a high negative predictive value in order to avoid unnecessary biopsies and surgical resections. However, in this present study, no statistically significant added benefit in diagnostic performance was achieved by adding SWE. This may have been due to the excellent diagnostic performance of the B-mode US category of high suspicion by itself, with a sensitivity of 95.1%, a specificity of 85.12%, a positive predictive value of 82.9%, a negative predictive value of 95.8%, and an accuracy of 89.5%.
There were several limitations of this study. First, selection bias was inevitable due to its retrospective design. Second, this study was based on results obtained by two operators, and we did not assess interobserver variability. However, SWE is known to be relatively reproducible and minimally operator-dependent. All thyroid nodule data were obtained using the same protocol to minimize operator bias. Third, this study included a relatively small number of thyroid nodules and only presented experiences from a single center. Therefore, our findings should be verified by a prospective study with a large population. Lastly, there was an unusually high proportion of malignant nodules in our study group. The patients included in the present study who had indeterminate or suspicious lesions underwent US FNA or surgical removal.
SWE is a valuable additional diagnostic tool for accurately stratifying the risk of thyroid nodules in routine clinical settings. A combination of B-mode US with SWE yielded improved diagnostic performance compared to B-mode US findings only, although the differences in diagnostic performance were statistically non-significant.

HIGHLIGHTS

▪ Age, a taller-than-wide shape, the presence of calcifications, and Emax were independent predictors of thyroid malignancy.
▪ The combination of B-mode ultrasonography with shear wave elastography (SWE) yielded improved diagnostic performance.
▪ SWE is a valuable additional diagnostic tool for predicting malignancy in thyroid nodules.

CONFLICT OF INTEREST

No potential conflict of interest relevant to this article was reported.

Notes

AUTHOR CONTRIBUTIONS

Conceptualization: YMS. Data curation: YMS, EKY. Formal analysis: SJY. Methodology: YMS, MS. Project administration: YMS. Visualization: EKY. Writing - original draft: EKY. Writing - review & editing: YMS, EJK, YGE, WSP.

Fig. 1.
A 53-year-old woman with surgically proven papillary thyroid carcinoma. B-mode ultrasound showed hypoechoic mass with irregular margin and microcalcifications which was assessed as a high suspicion. Shear wave elastography (SWE) displayed heterogenous color elasticity signal with peripheral high SWE (red). Measured elasticity parameters of Emean of 78.0 kPa, Emax of 82.7 kPa and Emin of 56.8 kPa were obtained.
ceo-2019-01235f1.jpg
Fig. 2.
A 36-year-old man with surgically proven papillary thyroid carcinoma. B-mode ultrasound showed mixed echoic mass with well-defined margin and oval shape which was assessed as an intermediate suspicion. Shear wave elastography (SWE) displayed relatively heterogenous color elasticity signal. Measured elasticity parameters of Emean of 40.0 kPa, Emax of 54.7 kPa and Emin of 10.3 kPa were obtained.
ceo-2019-01235f2.jpg
Fig. 3.
A 57-year-old woman with surgically proven nodular hyperplasia. B-mode ultrasound showed isoechoic mass with well-defined margin and oval shape which was assessed as a low suspicion. Shear wave elastography (SWE) displayed relatively homogenous color elasticity signal. Measured elasticity parameters of Emean of 39.8 kPa, Emax of 68.4 kPa and Emin of 11.0 kPa were obtained.
ceo-2019-01235f3.jpg
Table 1.
Comparison of clinicopathologic and radiologic factor for predicting malignancy
Variable Malignant (n=41) Benign (n=54) P-value
Age (yr) 47.6±14.5 55.5±13.9 0.009
Sex 0.240
 Male 13 11
 Female 28 43
Mean US nodule size (mm) 12.0±7.3 15.7±10.2 0.110
US finding
 Composition <0.001
  Solid 41 29
  Mixed 0 25
 Echogenicity <0.001
  Marked hypo 8 0
  Hypo 26 10
  Iso 3 20
  Hyper 4 24
 Margin <0.001
  Irregular/microlobulated 33 11
  Smooth 8 43
 Shape <0.001
  Taller than wide 24 10
  Wider than tall 17 44
 Calcification <0.001
  Yes 30 8
  No 11 46
US category <0.001
 Low suspicion 0 26
 Intermediate suspicion 2 20
 High suspicion 39 8
Elasticity parameter
 Emax (kPa) 67.0±45.1 51.8±45.0 0.030
 Emean (kPa) 52.7±36.3 38.0±30.9 0.027
 Emin (kPa) 32.6±27.4 22.0±18.5 0.036
 Eratio 5.3±6.9 4.1±5.1 0.340

Values are presented as mean±standard deviation.

US, ultrasound; Emax, maximum pixel value for a region of interest (ROI); Emean, mean value for a 2-mm diameter ROI placed in the stiffest region; Emin, minimum pixel value for a ROI, Eratio, the ratios of mean stiffness for the lesion-to-normal parenchyma.

Table 2.
Multivariate logistic regression analysis of clinical and radiologic factor for predicting malignancy
Variable OR (95% CI) P-value
Age 0.90 (0.82–0.99) 0.028
Composition (reference:mixed) 0 (0–0) 0.990
Echogenicity (reference:hyperechogenicity) 0 (0–0) 0.230
Margin (reference:smooth margin) 3.0 (0.42–21.4) 0.360
Shape (reference:wider than tall) 11.3 (1.23–114.6) 0.040
Calcification (reference:no calcification) 15.0 (1.49–150.2) 0.021
US category (reference:low suspicion) 0 (0–0) 0.280
Emax 1.22 (1.2–1.57) 0.021
Emean 0.84 (0.72–1.01) 0.070
Emin 1.10 (0.98–1.35) 0.058

OR, odds ratio; CI, confidence interval; Emax, maximum pixel value for a region of interest (ROI); Emean, mean value for a 2-mm diameter ROI placed in the stiffest region; Emin, minimum pixel value for a ROI.

Table 3.
Diagnostic performance of each B-mode US features and elasticity parameters with optimal cutoff value on SWE for predicting malignancy
Variable Cutoff Sensitivity (%) Specificity (%) PLR NLR AUC
Composition Solid 100 46.30 1.86 0 0.731
Echogenicity Hypoechogenic 82.93 81.48 4.48 0.21 0.838
Margin Irregular 80.49 79.63 3.95 0.25 0.801
Shape Taller than wide 58.54 81.48 3.16 0.51 0.700
Calcification Presence 73.17 85.19 4.94 0.31 0.808
US category High suspicion 95.12 85.19 6.42 0.06 0.913
Emean >54.9 36.59 87.04 2.82 0.73 0.631
Emax >47.1 53.66 70.37 1.81 0.66 0.633
Emin >40.4 31.71 88.89 2.85 0.77 0.613
Eratio >1.94 92.68 29.63 1.32 0.25 0.609

US, ultrasound; SWE, shear wave elastography; PLR, positive likelihood ratio; NLR, negative likelihood ratio; AUC, area under the receiver operating characteristic curve; Emean, mean value for a 2-mm diameter region of interest (ROI) placed in the stiffest region; Emax, maximum pixel value for a ROI; Emin, minimum pixel value for a ROI; Eratio, the ratios of mean stiffness for the lesion-to-normal parenchyma.

Table 4.
Comparison of diagnostic performance of US category and combined US category with SWE parameters
Variable Sensitivity Specificity PPV NPV Accuracy
US category 95.12 (39/41) 85.19 (46/54) 82.98 (39/47) 95.83 (46/48) 89.47 (85/95)
US category+Emax (combination) 100 (41/41) 85.19 (46/54) 83.67 (41/49) 100 (46/46) 91.58 (87/95)
US category+Emean (combination) 97.56 (40/41) 87.04 (47/54) 85.11 (40/47) 97.92 (47/48) 91.58 (87/95)
US category+Emin (combination) 97.56 (40/41) 88.89 (48/54) 86.96 (40/46) 97.96 (48/49) 92.63 (88/95)

P-value P-value P-value P-value P-value

US category vs. US category+Emax 0.47 1.0 0.85 0.49 0.80
US category vs. US category+Emean 1.0 0.99 0.99 0.99 0.80
US category vs. US category+Emin 1.0 0.77 0.81 0.98 0.61
US category vs. US category+Eratio 1.0 1.0 0.82 0.99 0.99

Values are presented as percent (number).

US, ultrasound; SWE, shear wave elastography; PPV, positive predictive value; NPV, negative predictive value; Emax, maximum pixel value for a region of interest (ROI); Emean, mean value for a 2-mm diameter ROI placed in the stiffest region; Emin, minimum pixel value for a ROI; Eratio, the ratios of mean stiffness for the lesion-to-normal parenchyma.

REFERENCES

1. Kim EK, Park CS, Chung WY, Oh KK, Kim DI, Lee JT, et al. New sonographic criteria for recommending fine-needle aspiration biopsy of nonpalpable solid nodules of the thyroid. AJR Am J Roentgenol. 2002 Mar;178(3):687-91.
crossref pmid
2. Ha EJ, Moon WJ, Na DG, Lee YH, Choi N, Kim SJ, et al. A multicenter prospective validation study for the Korean thyroid imaging reporting and data system in patients with thyroid nodules. Korean J Radiol. 2016 Sep-Oct;17(5):811-21.
crossref pmid pmc
3. Hoang JK, Lee WK, Lee M, Johnson D, Farrell S. US Features of thyroid malignancy: pearls and pitfalls. Radiographics. 2007 May-Jun;27(3):847-60.
crossref pmid
4. Iannuccilli JD, Cronan JJ, Monchik JM. Risk for malignancy of thyroid nodules as assessed by sonographic criteria: the need for biopsy. J Ultrasound Med. 2004 Nov;23(11):1455-64.
crossref pmid
5. Alexander EK. Approach to the patient with a cytologically indeterminate thyroid nodule. J Clin Endocrinol Metab. 2008 Nov;93(11):4175-82.
crossref pmid pdf
6. Castro MR, Gharib H. Thyroid fine-needle aspiration biopsy: progress, practice, and pitfalls. Endocr Pract. 2003 Mar-Apr;9(2):128-36.
crossref pmid pdf
7. Cosgrove D, Barr R, Bojunga J, Cantisani V, Chammas MC, Dighe M, et al. WFUMB guidelines and recommendations on the clinical use of ultrasound elastography. Part 4: thyroid. Ultrasound Med Biol. 2017 Jan;43(1):4-26.
crossref pmid
8. Ophir J, Cespedes I, Ponnekanti H, Yazdi Y, Li X. Elastography: a quantitative method for imaging the elasticity of biological tissues. Ultrason Imaging. 1991 Apr;13(2):111-34.
crossref pmid
9. Nell S, Kist JW, Debray TP, de Keizer B, van Oostenbrugge TJ, Borel Rinkes IH, et al. Qualitative elastography can replace thyroid nodule fine-needle aspiration in patients with soft thyroid nodules: a systematic review and meta-analysis. Eur J Radiol. 2015 Apr;84(4):652-61.
crossref pmid
10. Asteria C, Giovanardi A, Pizzocaro A, Cozzaglio L, Morabito A, Somalvico F, et al. US-elastography in the differential diagnosis of benign and malignant thyroid nodules. Thyroid. 2008 May;18(5):523-31.
crossref pmid
11. Monpeyssen H, Tramalloni J, Poiree S, Helenon O, Correas JM. Elastography of the thyroid. Diagn Interv Imaging. 2013 May;94(5):535-44.
crossref pmid
12. Tanter M, Bercoff J, Athanasiou A, Deffieux T, Gennisson JL, Montaldo G, et al. Quantitative assessment of breast lesion viscoelasticity: initial clinical results using supersonic shear imaging. Ultrasound Med Biol. 2008 Sep;34(9):1373-86.
crossref pmid
13. Athanasiou A, Tardivon A, Tanter M, Sigal-Zafrani B, Bercoff J, Deffieux T, et al. Breast lesions: quantitative elastography with supersonic shear imaging--preliminary results. Radiology. 2010 Jul;256(1):297-303.
crossref pmid
14. Liu BX, Xie XY, Liang JY, Zheng YL, Huang GL, Zhou LY, et al. Shear wave elastography versus real-time elastography on evaluation thyroid nodules: a preliminary study. Eur J Radiol. 2014 Jul;83(7):1135-43.
crossref pmid
15. Cantisani V, D’Andrea V, Biancari F, Medvedyeva O, Di Segni M, Olive M, et al. Prospective evaluation of multiparametric ultrasound and quantitative elastosonography in the differential diagnosis of benign and malignant thyroid nodules: preliminary experience. Eur J Radiol. 2012 Oct;81(10):2678-83.
crossref pmid
16. Shuzhen C. Comparison analysis between conventional ultrasonography and ultrasound elastography of thyroid nodules. Eur J Radiol. 2012 Aug;81(8):1806-11.
crossref pmid
17. Colakoglu B, Yildirim D, Alis D, Ucar G, Samanci C, Ustabasioglu FE, et al. Elastography in distinguishing benign from malignant thyroid nodules. J Clin Imaging Sci. 2016 Dec;6:51.
crossref pmid pmc
18. Park AY, Son EJ, Han K, Youk JH, Kim JA, Park CS. Shear wave elastography of thyroid nodules for the prediction of malignancy in a large scale study. Eur J Radiol. 2015 Mar;84(3):407-12.
crossref pmid
19. Veyrieres JB, Albarel F, Lombard JV, Berbis J, Sebag F, Oliver C, et al. A threshold value in Shear Wave elastography to rule out malignant thyroid nodules: a reality. Eur J Radiol. 2012 Dec;81(12):3965-72.
crossref pmid
20. Chen M, Zhang KQ, Xu YF, Zhang SM, Cao Y, Sun WQ. Shear wave elastography and contrast-enhanced ultrasonography in the diagnosis of thyroid malignant nodules. Mol Clin Oncol. 2016 Dec;5(6):724-30.
crossref pmid pmc
21. Liu Z, Jing H, Han X, Shao H, Sun YX, Wang QC, et al. Shear wave elastography combined with the thyroid imaging reporting and data system for malignancy risk stratification in thyroid nodules. Oncotarget. 2017 Jun;8(26):43406-16.
crossref pmid pmc
22. Haugen BR, Alexander EK, Bible KC, Doherty GM, Mandel SJ, Nikiforov YE, et al. 2015 American Thyroid Association management guidelines for adult patients with thyroid nodules and differentiated thyroid cancer: the American Thyroid Association guidelines task force on thyroid nodules and differentiated thyroid cancer. Thyroid. 2016 Jan;26(1):1-133.
crossref pmid pmc
23. Azizi G, Keller JM, Mayo ML, Piper K, Puett D, Earp KM, et al. Thyroid nodules and shear wave elastography: a new tool in thyroid cancer detection. Ultrasound Med Biol. 2015 Nov;41(11):2855-65.
crossref pmid
24. Kwak JY, Han KH, Yoon JH, Moon HJ, Son EJ, Park SH, et al. Thyroid imaging reporting and data system for US features of nodules: a step in establishing better stratification of cancer risk. Radiology. 2011 Sep;260(3):892-9.
crossref pmid
25. Na DG, Baek JH, Sung JY, Kim JH, Kim JK, Choi YJ, et al. Thyroid imaging reporting and data system risk stratification of thyroid nodules: categorization based on solidity and echogenicity. Thyroid. 2016 Apr;26(4):562-72.
crossref pmid
26. Shin JH, Baek JH, Chung J, Ha EJ, Kim JH, Lee YH, et al. Ultrasonography diagnosis and imaging-based management of thyroid nodules: revised Korean Society of Thyroid Radiology consensus statement and recommendations. Korean J Radiol. 2016 May-Jun;17(3):370-95.
crossref pmid pmc
27. Wang D, He YP, Zhang YF, Liu BJ, Zhao CK, Fu HJ, et al. The diagnostic performance of shear wave speed (SWS) imaging for thyroid nodules with elasticity modulus and SWS measurement. Oncotarget. 2017 Feb;8(8):13387-99.
crossref pmid pmc
28. Bhatia KS, Tong CS, Cho CC, Yuen EH, Lee YY, Ahuja AT. Shear wave elastography of thyroid nodules in routine clinical practice: preliminary observations and utility for detecting malignancy. Eur Radiol. 2012 Nov;22(11):2397-406.
crossref pmid pdf
TOOLS
Download PDF File  Download PDF File
PubReader  PubReader
ePub Link  ePub Link
XML Download  XML Download
Full text via DOI  Full text via DOI
Download Citation  Download Citation
  Print
Share:      
METRICS
10
Web of Science
8
Crossref
13
Scopus
4,773
View
169
Download
Related article
Editorial Office
Korean Society of Otorhinolaryngology-Head and Neck Surgery
103-307 Park Tower officetel, Yongsan-dong 5-ga, Yongsan-gu, Seoul 04385, Korea
TEL: +82-2-711-9091   FAX: +82-2-3487-6603   E-mail: editor.eceo@gmail.com
Copyright © Korean Society of Otorhinolaryngology-Head and Neck Surgery.                 Developed in M2PI
Close layer
prev next